What methods can be used to report ammonia in wastewater to the EPA?

ID de documento

ID de documento TE122

Fecha de publicación

Fecha de publicación 24/08/2020
Pregunta
What methods can be used to report ammonia in wastewater to the EPA?
Resumen
Nitrogen ammonia methods for EPA reporting in wastewater
Respuesta
Nitrogen ammonia can be measured using a spectrophotometer (or colorimeter), or an ammonia ion selective electrode (ISE). Provided below are accepted or equivalent United States Environmental Protection Agency (USEPA) methods for wastewater reporting.


Nitrogen-Ammonia Nessler Method 8038- USEPA accepted for wastewater analysis (distillation required), Method 350.2.
Nitrogen-Ammonia TNTplus Method 10205-Equivalent
Nitrogen Ammonia Ion Selective Electrode (ISE) method 10001 and 10002-  These procedures can be used for Standard Methods for the Examination of Water and Wastewater 4500-NH3 E for USEPA NPDES reporting.



As with any method used for compliance reporting, always consult with your local regulatory authority.