How are results with decimals exported through modbus with an EZ series analyzer?

ID de documento

ID de documento TE12608

Fecha de publicación

Fecha de publicación 14/06/2020
Pregunta
How are results with decimals exported through modbus with an EZ series analyzer?
Resumen
Explanation of how results with decimals are exported with EZ series analyzers.
Respuesta
The EZ series analyzers are not able to send decimals through modbus protocols. Instead the result would be multiplied by a factor, usually 100 for common 0-1 ppm applications. The outgoing signal would then need to be divided by the same factor on the receiving equipment in order to restore the result to it's normal form and resolution.